MagCellect Human Naive CD4+ T Cell Isolation Kit
R&D Systems, part of Bio-Techne | Catalog # MAGH115
Key Product Details
Assay Procedure
Refer to the product datasheet for complete product details.
Reagents Provided
The MagCellect Human CD4+ T Cell Isolation Kit (Catalog # MAGH102) contains the following reagents:
- MagCellect Human CD4+ T Cell Biotinylated Antibody Cocktail
- MagCellect Streptavidin Ferrofluid
- MagCellect Plus Buffer (10X)
This kit contains sufficient reagents to process 1 x 109 total cells.
Other Supplies Required
- Ficoll-Hypaque™
- Hemocytometer
- Centrifuge
- Human Erythrocyte Lysing Kit (Catalog # WL1000)
- MagCellect Magnet (Catalog # MAG997) or equivalent
- 12 x 75 mm (5 mL) polystyrene round-bottom tubes
- Sterile Pasteur pipettes or transfer pipettes
NOTE: Reaction incubations must be carried out at 2 to 8° C in a refrigerator and not in an ice bath to avoid excessively low temperatures that can slow the kinetics of the optimized reactions.
Procedure Overview
R&D Systems Protocol for MagCellect™ Human Naive CD4+ T Cell Isolation Kit (Catalog # MAGH115)
- Prepare a single cell suspension of mononuclear cells.
- Wash the cells once with excess PBS.
- Centrifuge the cells for 10 minutes at 200 x g.
- Decant the supernatant. If necessary, remove red blood cells using R&D Systems Human Erythrocyte Lysing Kit (Catalog # WL1000).
- Resuspend the cells in a small volume of cold 1X MagCellect Plus Buffer.
- Perform a cell count.
- Adjust the cell concentration to at least 50 x 106 cells/mL with cold 1X MagCellect Plus Buffer.
- Transfer 50 x 106 cells (1 mL) into a 5 mL polystyrene tube.
- Add 100 μL of MagCellect Human Naive CD4+ T Cell Biotinylated Antibody Cocktail.
- Mix the cell-antibody suspension and incubate at 2 °C to 8 °C for 15 minutes.
- Add 150 μL of MagCellect Streptavidin Ferrofluid to the cell suspension.
- Mix gently.
- Incubate at 2 °C to 8 °C for 15 minutes.
- Place the reaction tube in the MagCellect Magnet.
- Incubate for 6 minutes at room temperature (18 °C to 25 °C).
- Transfer the supernatant containing the naïve CD4+ T cells into a new 5 mL tube.
- Repeat the magnetic depletion with the new tube containing the recovered cells. The supernatant obtained at the end of this step is the final depleted cell fraction containing the desired enriched naïve CD4+ T cells.
Technical Hints
- If sterile cells are required following the cell selection, the entire procedure should be carried out in a laminar flow hood to maintain sterile conditions. Use sterile equipment when pipetting reagents that will be reused at a later date.
- Avoid antibody capping on cell surfaces and non-specific cell tagging by working fast, keeping cells and solutions cold through the use of pre-cooled solutions and by adhering to the incubation times and temperatures specified in the protocol. Increased temperature and prolonged incubation times may lead to non-specific cell labeling thus lowering cell purity and yield.
- When processing different numbers of cells observe the following guidelines: keep antibody cocktail and ferrofluid incubation times and temperatures the same; keep the cell density at 10 x 107 cells/mL; add 10 μL of the antibody cocktail per 1 x 107 cells being processed; add 12.5 μL of Streptavidin Ferrofluid per 1 x 107 cells being processed.
- When processing 20 x 107 cells or fewer, use the 12 x 75 mm (5 mL) tubes with the MagCellect Magnet horizontally positioned to accommodate up to six 5 mL tubes. Do not process more than 20 x 107 cells in each 5 mL tube and do not exceed a total reaction volume of 3 mL in each tube. A reaction volume of 2 mL is recommended for processing 10 x 107 cells. A reaction volume of 1 mL is recommended when processing 5 x 107 or fewer cells. Reaction volume adjustments must be made using 1X MagCellect Buffer just prior to the magnetic separation step.
- When processing greater than 20 x 107 cells, use the 17 x 100 mm (15 mL) tubes with the MagCellect magnet vertically positioned to accommodate up to two 15 mL tubes. Do not process more than 60 x 107 cells in each 15 mL tube and do not exceed a total reaction volume of 9 mL in each tube. When using this larger tube, increase the reaction volume before the magnetic separation step according to the following formula: 3 mL for each 20 x 107 cells processed. Also increase the magnetic incubation time described in to 8 minutes. Reaction volume adjustments must be made using 1X MagCellect Buffer just prior to the magnetic separation step.
Customer Reviews for MagCellect Human Naive CD4+ T Cell Isolation Kit