Minimum Sample Volume |
50 µL |
50 µL |
100 µL |
100 µL |
Sample Delivery |
Vacuum |
Electrokinetic |
Electrokinetic |
Vacuum |
Typical Separation Time |
6–10 min (molecule-dependent) |
Reduced IgG: 25 min, Non-reduced IgG: 35 min |
Reduced IgG: 5.5 min Non-reduced IgG: 8 min |
40 – 50 min (molecule-dependent) |
Detection Capability |
UV Absorbance at 280 nm Fluorescence: Ex 280 nm, Em 320– 450 nm |
UV Absorbance at 220 nm |
UV Absorbance at 220 nm |
Fluorescence: Ex 280 nm, Em 320-450nm |
Typical Voltage |
Pre-focusing: 1500 V, focusing: 3000 V |
Separation: 5750 V |
Separation: 4200 V |
Pre-focusing: 500 V and 1000 V Focusing: 1500 V |
Sample Injections per Cartridge |
100 guaranteed, 200 maximum (max 25 batches) |
100 guaranteed, 500 maximum (max 25 batches) |
100 guaranteed (max 25 batches) |
Maximum 15 injections |
Maximum Sample Injections per Batch |
100 |
48 |
96 |
1 (fractionation) 4 (cIEF) |
pI/Size Range |
2.85–10.45 |
10–270 kDa |
10–270 kDa |
3-10 |
pI/Sizing CV |
1% |
≤2% |
<2% |
1% |
CV for Peaks >10% Composition |
≤5% (Intra-batch), ≤6% (Inter-batch) |
N/A |
N/A |
≤10% (Inter-batch) |
Relative Migration Time CV |
N/A |
<1% for reduced IgG |
<5% |
N/A |
pI/Sizing Resolution |
0.05 pI units (for wide range 3–10 ampholytes) |
≥1.5 for NGHC/HC IgG Standard |
≥1.0 for NGHC/HC IgG Standard |
N/A |
Dynamic Range |
2 logs |
2 logs |
2 logs |
N/A |
Linearity |
>0.995 |
>0.995 |
>0.995 |
N/A |
Sensitivity (LOD) |
0.7 µg/mL (Native fluorescence) 3.0 µg / mL(Absorbance) (Values based on a monoclonal antibody) |
0.3 µg/mL (Value based on Internal Standard) |
0.6 µg/mL (Value based on Internal Standard) |
N/A |
Sample Tray Options |
96-well plates or 48 vials |
96-well plates only |
Power |
100 V–240 V (AC), 50/60 Hz, 500 W |
Voltage Range |
0–6,500 V |
Temperature Control Range |
4-25 °C |
10-25 °C |
Dimensions |
44cm H x 42cm W x 61cm D |
Weight |
46 kg (100 lb) |